Published on July 13, 2026 at 03:04 AM
A hospital facility technician is tasked with mapping the 5-gauss safety zone around a newly installed MRI scanner. Within this zone, no ferromagnetic objects (metal trolleys, wheelchairs, oxygen cylinders) may be stored, as they could be pulled into the MRI bore with dangerous force. The boundary zone must be precisely identified and marked.
Manual testing with magnets and rolling carts is dangerous and impractical. A magnetic field meter such as the Extech SDL900 with a Hall sensor probe makes mapping possible: the technician walks slowly through the room with the probe, measures the field strength at each position (the datalogger stores the readings automatically), and draws the 5-gauss boundary on a floor plan.
Hall Sensor Physics
The Hall effect (Edwin Hall, 1879): when a current-carrying material is exposed to a magnetic field, a voltage difference is produced perpendicular to the current flow. This voltage is proportional to the magnetic field strength.
Modern Hall sensors use semiconductor material (GaAs or InSb). They output a millivolt signal proportional to the applied magnetic field:
- At 1 mT (10 gauss): approx. 1-10 mV output signal.
- The instrument electronics amplify and digitize the signal.
- Display in gauss (G), tesla (T) or milligauss (mG).
Conversion: 1 mT = 10 G. 1 μT = 10 mG. The Earth’s magnetic field is approx. 0.25-0.65 G depending on geographic location.
AC vs. DC Magnetic Fields
The SDL900 measures BOTH field types:
DC magnetic field: Static, constant direction. Generated by:
- Permanent magnets (neodymium, ferrite).
- Electromagnets with DC current.
- The MRI main magnetic field (0.5-3 tesla at these field strengths).
- The Earth’s magnetic field.
- Motors at standstill (rotor magnet).
AC magnetic field: Time-varying, oscillating. Generated by:
- Mains power lines (50 Hz).
- Transformers.
- Electric motors in operation.
- Induction cookers.
- Switch-mode power supplies.
- Radio and wireless transmitters (higher frequencies).
For DC measurement, the instrument uses the ZERO button for zeroing, which compensates for the Earth’s magnetic field and reference drift.
N/S Pole Indication
In DC mode, the instrument displays not only the magnetic field strength but also the POLARITY, showing whether the sensor is facing a north or south pole. This is important for:
- Permanent magnet calibration: verifying correct polarity after assembly.
- Motor bearing analysis: checking rotor magnet polarity.
- Compass and navigation calibration.
- Scientific experiments with a defined field direction.
For AC fields, polarity indication is irrelevant (the field alternates 100 times per second with 50 Hz mains power anyway).
Automatic Temperature Compensation
Hall sensors have a temperature-dependent output sensitivity. They respond differently at -20 °C than at +40 °C. Without compensation, the displayed field strength can deviate by 10-15 % across the operating temperature range.
The SDL900 has an internal temperature sensor at the Hall element, and the microprocessor corrects automatically. The result: consistent readings from 0-50 °C without manual calibration.
Datalogger on SD Card
The SDL900 saves readings continuously to an SD card in Excel format. This is useful for:
- Safety zone mapping: The technician walks through the room with the probe, and the datalogger stores position + field strength. Excel analysis follows later.
- Long-term monitoring: In factories with suspected EMI interference, 24-hour logging shows when and where peaks occur.
- Trend analysis: Comparison with reference measurements from previous months.
Configurable sampling rate of 1-3600 seconds. At a 10-second sampling interval, 24 hours = 8,640 data points, which is practically unlimited on a 2 GB SD card.
Applications
Hospital MRI safety: 5-gauss boundary zone mapping. Regular verification.
Electronics manufacturing: MEMS sensors, Hall sensor production — testing the ambient magnetic field for calibration.
Permanent magnet manufacturers: QC in the production of neodymium magnets. Verification of field strength after magnetization.
EMI diagnostics: For interference in electronic systems — locating the source of the disturbance.
Electric motor maintenance: Rotor magnet wear, stator field inspection.
Semiconductor cleanrooms: Some processes are sensitive to magnetic fields. Baseline measurement + continuous monitoring.
Scientific research: Universities, research laboratories. Physics lab courses.
Aerospace calibration: Compass testing, sensor testing.
Military and defense: Magnetic signature analysis.
Archaeological excavation: Fluxgate magnetometers for metallic objects in the ground.
Limitations
- Single-axis probe: Measures only one field component. For a full 3D vector measurement, 3 measurements must be taken with the probe rotated.
- AC frequency range: Effective up to approx. 500 Hz. Higher frequencies (RF, kHz) require special RF probes.
- Not for high-field applications (inside an MRI): The sensors are calibrated for ambient field measurements (up to approx. 200 mT). Inside the MRI bore (1.5-3 T), the sensor saturates.
- Not for precision measurement of the Earth’s magnetic field: Geophysical applications require fluxgate or proton magnetometers.
What You Get for Your Money
Extech SDL900 AC/DC magnetic field meter with datalogger, Hall sensor with automatic temperature compensation, N/S pole indication for DC measurement, Zero button, Data Hold, MIN/MAX function, configurable sampling rate of 1-3600 seconds, SD card datalogging in Excel format (SD card included), 99 manual reading memory locations, RS-232 interface, single-axis magnetic probe with protective cover, 6 AA batteries, hard carrying case. 846 EUR.
A standard instrument for MRI safety, EMI diagnostics, permanent magnet QC and scientific research, wherever AC and DC magnetic field measurements with data logging are required. Combining two field types in one instrument saves investment compared with separate single-purpose devices.